Description
Product Description
Features
- The NEMA IEC Body Phantom Set™ consists of a body phantom, a lung insert and an insert with six spheres with various sizes
- It is designed in accordance with the recommendations by the International
Electrotechnical Commission (IEC) and modified by the National Electrical manufacturers Association (NEMA) - It is recommended for use in the evaluation of reconstructed image quality in whole body PET imaging
Applications
- Simulation of whole-body imaging especially using PET and camera-based coincidence imaging techniques
- Evaluation of reconstructed image quality in wholebody PET and camera-based coincidence imaging
- Determination of the coincidence count rate characteristics in brain and cardiac imaging
- Evaluation of the relationship between true coincidence count rate and radioactivity
- Determination of the address errors caused by address pile up
- Evaluation of the count loss correction scheme
- Research
Item Measurements
- Interior length of phantom: 180 mm
- Fillable spheres (6) inner diameter: 10 mm, 13 mm, 17 mm, 22 mm, 28 mm, and 37 mm.
- Distance from sphere plane to inside wall: 70 mm
- Volume of empty cylinder: 9.7 liters
- Cylindrical insert dimension: Outside diameter: 51 mm; Length: 180 mm
Product Code
Product code | Description |
|---|---|
NM-5250 - 0161 | NEMA IEC Body Phantom Set™ |
* International Standard: Radionuclide imaging devices –Characteristics and test conditions – Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC), 61675-1, Geneva, Switzerland, 1998.
* Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2001.
PP-MIP-AU-0029-2