Features

Rotem RAM GENE-1 mark II

The RAM GENE-1 mark II,designed for work in extreme conditions, features a ruggedized casing and backlight and contains a specially designed energy filter embedded in the cap for dose rate radiation monitoring according to Ambient Dose Equivalent, with the additional flexibility of a contamination meter.

The RAM GENE-1 is equipped with a recessed, internal GM pancake detector using a thin mica window protected by a stainless steel screen.

The transition between measuring dose rate radiation and contamination is achieved by simply removing or placing the plastic cap.
 

Features

  • Large, four digit, continuous ranging, LCD display
  • Simple, three key operation
  • Large 1.75" pancake GM detector
  • Suitable for alpha, beta or gamma contamination detection
  • Unit readouts of count rate (CPS in contamination mode and (µSv/hr) in dose mode.
  • Audible and visual (LED) count rate indication
  • Smart software provides fast, yet stable response
  • Dead time correction
  • Alarms for overflow, low battery and detector failure
  • Automatic self diagnostic routines
  • Durable case with splash resistant keypad
     

Specifications

The RAM GENE-1 is equipped with a recessed, internal GM pancake detector using a thin mica window protected by a stainless steel screen. The transition between measuring dose rate radiation and contamination is achieved by simply removing or placing the plastic cap.
 

Technical Information

Detector: 1.75″ Pancake GM tube – LND73118 or equivalent

Count Range: 0 – 42,000 cps

Measuring Range: 0.05 µSv/hr to 7 mSv/hr (5 µR/hr to 700 mR/hr)

Sensitivity: Approximately 5.8 cps/µSv/hr (58 cps/mR/hr) – 137Cs
 

Gene 1

gene1-1
gene1-2

Downloads

PP-MIP-AU-0029-2
Date of preparation: July 2026